GET THE APP

Reverse Phase - High-performance liquid chromatography method for simultaneous estimation of levocetirizine and pseudoephedrine in raw and tablet formulation | Abstract
Scholars Research Library

Scholars Research Library

A-Z Journals

+44 7389645282

Der Pharmacia Lettre

Abstract

Reverse Phase - High-performance liquid chromatography method for simultaneous estimation of levocetirizine and pseudoephedrine in raw and tablet formulation

Author(s): Gunasekar Manoharan, Mohammed Al Bratty and Amnh Abdulrahman Mohammed Abdulhaq

A simple, reproducible method was developed for the determination of levocetirizine and pseudoephedrine in tablet formulation using reverse phase HPLC method. HPLC separation was carried over on a Thermo hypersil – keystone C18 (250 x 4.6mm, 0.5m) isocratic mode column using a mobile phase comprising a 50:50 (% v/v) of acetonitrile and 0.5mM phosphate buffer. The detection was proceeded out by UV detector at 257 nm. The range taken for linearity for levocetirizine and pseudoephedrine were 5 – 25 mg/ml and 120 – 600 mg/ml. The percentage recovery was 100.11 % and 110.04 % for levocetirizine and pseudoephedrine respectively. The amount of levocetirizine and pseudoephedrine found per tablet was 5.05 mg and 119.84 mg respectively.