GET THE APP

Development and Validation of High Performance Liquid Chromatography (HPLC) method for the Determination of Bosantan in Pure and Formulated Forms. | Abstract
Scholars Research Library

Scholars Research Library

A-Z Journals

+44 7389645282

Der Pharmacia Lettre

Abstract

Development and Validation of High Performance Liquid Chromatography (HPLC) method for the Determination of Bosantan in Pure and Formulated Forms.

Author(s): K. Siddappa and Prashant C. Hanamshetty

A reversed-phase high performance liquid chromatographic (RP-HPLC) method for the determination of Bosantan (BOS) in bulk material and pharmaceutical formulations has been developed and validated. The chromatographic separation of BOS was performed on an Inertsil C8 column (5 m particle size, 15 cm length x 4.6 mm i.d) with CLC ODS (4 cm x 4.6 mm, i.d.) as a guard column to protect analytical column, a mobile phase consisting of acetate buffer (pH 5.5) acetonitrile in the ratio of 20:80 (v/v) at a flow rate of 1.0 ml/min at ambient temperature using UV detector at 230 nm. The developed method gave good retention time of 2.225 min. The method was found to be specific to BOS and able to resolve the drug peak from formulation excipients. The calibration curve was noticed to be linear over the concentration range of 5-150 μg/ml (R2 = 0.9999). The method was validated statistically for its selectivity, linearity, precision, accuracy and robustness. In view of this, the method could be employed as a more convenient and efficient option for the analysis of BOS in drug substance and formulations.